15. Integrations with Other Tools¶
This chapter answers questions about using UniPlot together with other software and data sources: Python, MATLAB, databases, ASAM ODS servers, MORPHEE, TOPexpert, INCA and the emission and RDE add-ins.
15.1. Python, COM, DLLs and Databases¶
15.1.1. Can I use Python with UniPlot?¶
Python is not built into UniPlot. Scripting and formulas in UniPlot use UniScript (the language is similar to MATLAB) or Lua. You can still combine UniPlot with Python in several ways:
Start a Python script from UniScript with system or ShellExecute and exchange data through files (for example NC2, CSV or MAT files) or the command line.
Control UniPlot from other programs through its COM interface (see UniScript as an COM-Server) or through DDE (see Overview DDE). From UniScript you can access other COM servers with GetObject.
Write a DLL in C or C++ with a plain C interface that uses only standard data types (
double,int,long, …) and register its functions with RegisterFunction.Start UniPlot or the stand-alone UniScript interpreter
us.exefrom Python on the command line (see UniPlot-Tools).
See also: How can I write a DLL and call one of the DLL functions from UniScript?, How can I read NC2 files with Python?.
15.1.2. How can I read NC2 files with Python?¶
The simplest way is to convert the NC2 file to a text file in CDL format with the UniPlot tool ncdump.exe and to parse that text file in Python:
ncdump.exe measurement.nc2 > measurement.cdl
The reverse conversion is done with ncgen.exe. Both tools are described in ncdump.exe and ncgen.exe.
Alternatively, export the data from UniPlot to a format that Python reads directly (for example CSV, Excel or MAT) with File=>Export Data. An NC2 software development kit with a Python example that loads an NC2 file into a pandas DataFrame is also available from UniPlot support.
See also: Overview UniPlot Data Files.
15.1.3. Can UniPlot connect to a database (SQLite, MySQL, PostgreSQL, Oracle, Microsoft SQL Server)?¶
UniPlot has the following database interfaces out of the box:
SQLite through the UniScript functions sqlite3_open, sqlite3_execute and sqlite3_close (see SQLite (SQL)).
MySQL and MariaDB through the UniScript functions mysql_connect, mysql_query and mysql_fetch_rows (see MySQL, MariaDB).
PostgreSQL through the UniScript functions pq_connectdb, pq_exec and pq_getvalue (see Overview PostgreSQL).
ASAM ODS servers through the ODS add-ins (see Which ASAM ODS versions and interfaces does UniPlot support? Are encrypted (SSL) connections possible?). ASAM ODS servers can themselves run on top of databases such as Oracle, so such data can be reached through the ODS interface.
There is no built-in connection to Microsoft SQL Server or Oracle. A workaround is to export the data from the database to a file format that UniPlot imports (for example CSV, Excel or MDF4) and to import the file as usual.
See also: Importing Data.
15.1.5. How can I fill an existing Excel file via COM from UniScript, and why is it slow?¶
Use the functions in the script file excel_com.ic in the folder <uniplot>\script\io. They control Microsoft Excel through its COM interface, so Excel must be installed. Copy the file into the folder with your own .ic files and load it.
ExcelWrite_COM opens an existing workbook, writes a matrix into a sheet and saves the file:
smData = ["Speed", "Torque"; "1000", "120"; "2000", "180"];
smType = ["S", "S"; "", ""; "", ""]; // "S" = text, "" = number
ExcelWrite_COM(["d:\\Results\\Report.xlsx", "Sheet1", "appendtosheet"], smData, smType, [3, 2]);
"appendtosheet"writes into the existing sheet. If the sheet does not exist, a new sheet is added."appendsheet"always adds a new sheet.The last argument is the start cell as
[row, column].[3, 2]is cell B3.With
"appendtosheet"and"appendsheet"the workbook is saved in the opened file (Save()). Only the type"new"creates a new file (SaveAs()) and overwrites an existing file with that name.
If you write your own COM code, use excel_com.ic as the example: ox = GetObject("Excel.Application"), ox.Workbooks.Open(ssFile), activate the sheet, write the data with as.Range("B3:F20").Value = vmData, then call ox.ActiveWorkbook.Save() and ox.ActiveWorkbook.Close().
To make writing fast:
Collect all values in one matrix and write the whole range with one
Range(...).Valueassignment. Do not write cell by cell.Set
ox.Application.ScreenUpdating = 0while you write, and set it back to1at the end.
See also: How can I write data into an Excel file or an existing Excel template with a script?, Overview Excel Files, GetObject.
15.2. MATLAB and Simulink¶
15.2.1. How can I convert NC or NC2 files to MAT, CSV or Excel files, also automatically with us.exe?¶
In UniPlot choose File=>Export Data, open the file to export and select the target format (for example MATLAB .mat, text/CSV or Excel). This also works for very large files. A MATLAB installation is not required. See Export of netCDF files to MATLAB.
For automatic conversion without user interaction, run a script that calls NC_ExportData with the stand-alone interpreter us.exe (see us.exe and UniScript as a Stand-Alone Program):
Write the conversion into a script file, for example
export.ic. Double the backslashes in UniScript strings ("C:\\Data\\file.nc").To pass file names to the script, read them in the script with getargs:
args[1]isus.exe,args[2]is the script andargs[3]is your first own argument.Start
us.exewith the script file as the first argument. Put paths that contain spaces in double quotes:"C:\Program Files\UniPlot Software\<release>\Program\us.exe" "D:\Scripts\export.ic" "D:\Data\file.nc2"
Check the exit code:
0means success.us.exewrites a log fileuniscript-<id>.loginto the log folder of the UniPlot user directory.
From MATLAB, start the command with [status, cmdout] = system(command).
Why does us.exe return 1 and write no file? us.exe does not execute UniScript statements given on the command line. It takes the first argument as the name of a script file. A call such as us.exe svExportNames = NC_ExportData(...) therefore cannot load a script, ends with exit code 1 and writes no file.
See also: Import of MATLAB Data Files.
15.2.2. Why can MATLAB not load a .mat file exported from UniPlot?¶
MATLAB cannot read some UniPlot-specific content of the exported file:
global attributes whose names start with an underscore (
_),channels with text content that is not valid UTF-8 (often comment channels),
the variable
channel_infothat UniPlot generates to store the channel attributes.
Workaround:
In the NC editor remove all global attributes that start with
_.Export only channels that contain no non-UTF-8 text.
In MATLAB load the file without
channel_info:
vars = who('-file','file.mat');
vars = vars(~strcmp(vars,'channel_info'));
load('file.mat', vars{:})
See also: Import of MATLAB Data Files.
15.2.3. Why can UniPlot not load MAT files with enumeration signals (Simulink Data Inspector, To File block)?¶
MAT files that contain enumeration signals do not fully follow the MATLAB file standard, so the UniPlot MATLAB import filter cannot load them. Use another logging or export format for such signals, for example MDF4.
See also: Import of MATLAB Data Files.
15.3. ASAM ODS Servers¶
15.3.1. Which ASAM ODS versions and interfaces does UniPlot support? Are encrypted (SSL) connections possible?¶
UniPlot has two ODS add-ins:
ASAM-ODS Browser (RPC) for servers with the RPC interface (ODS API 3), see ASAM-ODS Browser (RPC). The menu entries are AODS-3 Browser and AODS-3 Change Login.
ASAM-ODS Browser (CORBA) for servers with the CORBA interface (ODS 4 and 5), see ASAM-ODS Browser (CORBA).
The ODS HTTP API of ASAM ODS 6 is not supported in the standard product.
SSL/TLS-encrypted connections are not supported either. The ASAM-ODS Browser (CORBA) connects with plain CORBA/IIOP. If your IT policy requires encryption, provide it on the network level, for example with a VPN connection between client and server.
15.3.2. Connecting to an ASAM-ODS RPC server fails (user or password error, AODS-3 Configuration does not work)¶
Check the following points in this order:
The connection data in File > AODS-3 Change Login: the computer name or IP number, the RPC number, the user name and the password. A wrong RPC number is the most frequent cause of failed connections (see Login).
The server-specific options in Tools > AODS-3 Configuration (see Configuration).
If the same server works with another ODS client and the login still fails after the server was upgraded or reconfigured, check the server itself (server version, user and project configuration). Login errors can be caused by the server configuration and are independent of UniPlot.
Use a current UniPlot release. Problems with the AODS-3 configuration dialog in some older releases are fixed in later releases.
See also: ASAM-ODS Browser (RPC).
15.3.3. Some projects, test records or measurements of an ODS server are not visible in the ODS browser¶
Fixed in R2020.4
Update to R2020.4 or later. Older releases had two errors with the 64-bit instance IDs of ASAM ODS objects in the ASAM-ODS Browser (CORBA):
Before R2020.3, ID values larger than 2,147,483,647 were not handled correctly. Child objects of such entries (for example the test records of a project or the matrices of a test record) could not be retrieved. Whether an entry was visible depended on its numeric ID, not on its content.
Before R2020.4, a node whose 64-bit instance address had a high part other than 0 could not be opened in the navigation tree.
If entries are still missing in a current release, the cause is probably on the server side or in the data model. Check the server with another ODS client. Then send support the UniPlot version, the ODS server type and version, and an example of a project that is not shown.
See also: ASAM-ODS Browser (CORBA).
15.3.4. I cannot connect to the ASAM ODS (CORBA) server from home office through VPN or Zscaler. What can I do?¶
Enter the server in File > AODS-5 Change Login in a form that your VPN or proxy can reach. UniPlot uses exactly the computer name that you enter there:
Enter the fully qualified host name (FQDN) of the name service host if your VPN or proxy (for example Zscaler) works with DNS names.
Otherwise enter the IP address.
Also enter the port (standard 2809) and the binding name. UniPlot builds the CORBA address corbaloc:iiop:1.2@<computer>:<port>/NameService from these entries.
If the name service is reachable but the connection still fails: after the name service lookup, UniPlot connects to the address that the ODS server publishes in its object references. If the server publishes an internal IP address or a short host name that cannot be reached through the VPN, the connection fails. This must be solved in the network or server configuration by your IT or by the supplier of the ODS server.
See also: ASAM-ODS Browser (CORBA), Which ASAM ODS versions and interfaces does UniPlot support? Are encrypted (SSL) connections possible?.
15.3.5. Deleting a saved connection in AODS-5 Change Login fails with “_AODS_4_LogInDialog_DeleteConnection: undefined variable”¶
Fixed in R2026.4
This was an error in the login dialog of the ASAM-ODS Browser (CORBA). Deleting a saved connection setting failed with this message. Update to R2026.4 or later.
See also: ASAM-ODS Browser (CORBA).
15.3.6. The ODS3 browser download hangs at 100% when a key contains two measurements¶
The download of a measurement from the ODS3 browser could hang at 100% in some releases if two measurements were stored below one key. The problem is corrected in a later release; update UniPlot to the current release.
See also: ASAM-ODS Browser (RPC).
15.3.7. How can I export the table shown in the ASAM-ODS import dialog to Excel?¶
The list in the import dialog cannot be exported directly. Use two steps instead:
Import the ASAM-ODS data into UniPlot so that it is saved as an NC/NC2 file.
Open that file and choose File=>Export Data to write an Excel file.
See also: ASAM-ODS Browser (RPC).
15.3.8. The filter does not work on an NC file created by the ODS browser. Can the ODS browser create NC2 files?¶
The ODS browser always writes NC files, not NC2 files. If the data point filter does not work on such a file, specific channels of the file are the cause. Delete the channels named in the error message from the NC file (for example in the NC editor) and apply the filter again.
15.3.9. How do I get the ASAM path of a measurement from its application element and instance IDs (UniScript)?¶
Do not build the path yourself. Get the instance element for the two IDs and ask it for its path. The result is the complete ASAM path, including the parent test.
// aop: ODS session object of your script
// aid: ID of the application element, iid: ID of the instance
// 64-bit IDs are given as [low, high]; for small IDs high = 0
ieObj = _AODS4_GetInstanceById(aop, [aid, 0, iid, 0]);
if (ieObj != 0) {
ssPath = _AODS4_InstanceElement_GetAsamPath(ieObj);
}
If you already have the application element object
aeObj, call_AODS4_GetInstanceById(aop, aeObj, [iid, 0])instead.ieObj.getAsamPath()returns the path directly from the server.The reverse direction is
aop.aoApplStructure.getInstanceByAsamPath(ssPath)._AODS4_MakeAsamPathonly joins path pieces that you already know. It cannot find the parent test.
See also: ASAM-ODS Browser (CORBA).
15.4. MORPHEE and TOPexpert¶
15.4.1. How do I import MORPHEE test results into UniPlot?¶
Enable the add-in MORPHEE import (v1.0) in Tools=>Add-In Manager.
Choose File=>Import Data and select the type
MORPHEE(*.files).Select the
*.filesfile of the result folder. The add-in builds the channel structure from the list of result files in this file.
For single MORPHEE files the add-in also offers the types MORPHEE XPAR File (*.XPAR), MORPHEE Log File (*.MLOG), MORPHEE MTABLE File (*.MTABLE), MORPHEE CSV File (*.CSV) and MORPHEE MNPT File (*.TXT).
See also: Import of MORPHEE files.
15.4.2. The MORPHEE import does not import the first value of the first column and the last column of an .mtable file¶
This error of the MORPHEE import add-in in earlier R2026 releases is corrected in R2026.5 (improved MTABLE import). After you install R2026.5, delete a file mtable_import_filter.ic that was installed as an interim fix from the autoload folder of the user directory. Otherwise it overrides the corrected official import. The user directory can be opened with Help=>Open User Directory.
See also: Import of MORPHEE files.
15.4.3. How can I open the UUT CSV file created by MORPHEE in UniPlot?¶
The values in the MORPHEE UUT file are stored as text in quotation marks, so UniPlot does not recognise them as numbers during the import. Configure the data format of the MORPHEE test so that the quotation marks (") are no longer written; then the CSV file can be imported as usual. A UniScript conversion script that turns such a CSV file into the UTX format is available from UniPlot support; copy it into the autoload folder of the user directory and restart UniPlot.
See also: Importing Data.
15.4.4. How can I load a MORPHEE database backup (.bak) into a newer SQL Server version?¶
Do not restore the backup with SQL Server tools. MORPHEE provides the programs ExportMorpheeBench.exe (export on the old system) and CreateMorpheeBench.exe (import on the new system) in the BIN directory of its installation. See the MORPHEE documentation or the MORPHEE support for details.
15.4.5. MORPHEE starts UniPlot with a sequence, but the document cannot be loaded (UAC virtualization, no points in file)¶
If UniPlot is started from a process that runs elevated (as administrator), it cannot load files that were saved by the current user. Start UniPlot in the session of the current user instead, for example with the MORPHEE process control (StartProcess with the option to start as the current user) or from a .bat file that is executed by the user.
The command line for a Sequencer sequence is:
uniplot.exe /ws "<workspace folder>" /seq "<sequence>" /rv "DATAFILEPATH=<file>"
Paths must end with the folder name without a trailing backslash. A UniScript file can be started at program start with the command line switch /StartupFileName (/Ini is the same switch); a script can read the command line with AppGetCommandLine. See also Can I start UniPlot with command-line arguments?.
See also: How does the TOPexpert “UniPlot integration” work and why is the data in my document not updated?, How do I start a Sequencer sequence from the command line and pass values to it?.
15.4.6. How does the TOPexpert “UniPlot integration” work and why is the data in my document not updated?¶
TOPexpert Generic Evaluation starts UniPlot in automation mode and runs a Sequencer sequence (for example a picture export) with a command line of the form:
uniplot.exe /automation /ws "<workspace>" /seq "<sequence>" /rv "DOCPATH = ..., TEXTFILEPATH = ..., EXPORTPATH = ..."
The sequence uses the passed runtime values (see Runtime Values) to find the document and the data. If the exported pictures show the old plots stored in the IPZ document and the data is not refreshed, the runtime value that contains the data path is probably missing in the call. In the sequence, check with the action Snippet (functionality LOG RVs) which runtime values arrive.
See also: The Sequencer, Sequencer Tutorial.
15.4.7. An MF4 file exported from TOPexpert cannot be opened or plotted in UniPlot¶
Large MF4 files that TOPexpert creates from many measurement files (for example robustness evaluations) can be too large for UniPlot or can contain channels that block the import. Try the following:
Reduce the number of measurement files that are merged into one MF4 file.
Delete the channels that are named in the error message from the exported file and open it again.
Import the MF4 file directly into UniPlot (File=>Import Data) instead of exchanging it through the TOPexpert integration, and use a current UniPlot release.
See also: Import of MDF4 Files.
15.4.8. The TOPexpert merging tool reports licence problems. How can I check whether Measurement Processing licences are available?¶
If the merging tool cannot start, the licence server may not have a free licence of the Measurement Processing module at that moment. Open the LM-X client tool from the Windows Start menu (LM-X End-user Tools, LM-X License Server Client), search for the affected user and check which licences and modules are in use. If you need help, send the licence server log file to UniPlot support.
15.5. INCA, CAN and Other Import Add-ins¶
15.5.1. How can I read DCM or PACO calibration data files exported from INCA?¶
Use the add-in for calibration data:
Enable Calibration Data import (v1.0) in Tools=>Add-In Manager.
Choose File=>Import Data, select the type
Calibration Data File (*.dcm;*.DCM;*.xml)and select the files. PaCo files are XML files, so select them with the extension.xml.To plot maps and curves, select the dataset type
Calibration Plot (1D, 2D, 3D)in the Data Browser.Calibration Table (1D, 2D, 3D)shows the values as a table.
DCM files in version 2.0 are supported (version 1.0 is not).
See also: Import of Calibration Data Files.
15.5.2. A dialog of a third-party evaluation add-in (for example a channel assignment dialog) does not behave correctly¶
Dialogs and evaluations that are not part of the standard UniPlot distribution come from external add-ins that were developed by other vendors or for specific projects. UniPlot cannot change their behaviour. Contact the provider of the add-in. The add-ins that are installed on your computer are listed in Tools=>Add-In Manager.
15.5.3. The data import in the OCT (One Click Tool) fails with an error in startup.ic or “Visual Basic runtime error 440”¶
The scripts delivered with the OCT can fail when a UniPlot script, the add-in settings or the user directory cause a problem. Check the following:
Empty the
autoloadfolder of the user directory and restart UniPlot. If the error disappears, put the scripts back one by one to find the one that causes it.Check that the user home directory is writable. It must not be on a drive that is offline or write-protected. If necessary change the path in the registry key
HKEY_CURRENT_USER\Software\RSB\UniPlot5\settings(valueUser-Home-Directory).Check the OCT settings and the add-in settings, and any personal script that is loaded at startup.
See also: Help=>Open User Directory.
15.5.4. Formulas on an imported MF4 file stopped working: the time channels no longer have a device suffix¶
In current releases the device name suffix (for example _ETK_1) is deliberately not appended to the time channel names of imported MF4 files. Formulas and templates that refer to time channel names with a suffix must be changed to the new names. Calculated channels that are not supported by the import are not imported either.
See also: Import of MDF4 Files, Formula Configuration.
15.5.5. How do I import a CAN log (BLF or ASC) and select the DBC file?¶
The CAN add-in imports BLF and ASC log files. It needs a CAN database (DBC file) and imports only the signals that are defined in the DBC file.
Enable CAN Import (BLF, DBC, v0.2 beta-1) in Tools=>Add-In Manager.
Choose Tools > CAN Configuration (DBC file)… and select one or more DBC files (see Can-Configuration). The file dialog also accepts ARXML files.
Choose File=>Import Data, select the type
Vector CAN Log (*.blf; *.asc)and select the log files.
If no valid DBC file is set, UniPlot shows the message No DBC file selected: Goto Tools=>CAN Configuration to select a DBC file. Select the DBC file as in step 2.
The add-in also offers the types A&D CAN Log (*.tdfa) and G.i.N CAN Log (Data*F*; Data*F*.gz).
If the DBC dialog does not work in R2020.1 or older, update UniPlot to R2020.2 or later. Bus numbers for each DBC file can be set from R2020.4.
See also: Import of CAN Files.
15.5.6. How do I import PNRF files, and why does the PNRF import fail with an error?¶
The add-in does not read PNRF files itself. It uses the PNRF Reader of the manufacturer HBM, which must be installed first. If the reader is missing, the import fails.
Install the PNRF Reader. The setup is available from www.hbm.com.
Open Tools=>Add-In Manager, tick PNRF Import, click OK and restart UniPlot.
Choose File=>Import Data, select the type
PNRF File (*.pnrf)and select the files.
If the add-in causes errors in functions that do not use it (for example synchronising files), untick it. PNRF import is then not available.
See also: Import of PNRF Data Files.
15.5.7. The CONVERGE add-in is not visible in my UniPlot installation¶
The CONVERGE add-in is part of the installation from R2025.3 (CONVERGE import was introduced in R2025.3). Update UniPlot, then enable the add-in in Tools=>Add-In Manager. For older installations UniPlot support provides the add-in files (the folder rs_converge and its add-in file): unzip them into the addin folder of the installation, restart UniPlot and enable the add-in in the Add-In Manager.
See also: Import of CONVERGE files.
15.5.8. Can UniPlot read FEVIS data?¶
Not in current releases. The import of the FEVIS add-in is disabled: the add-in does not add a FEVIS file type to File=>Import Data. Export the data from the FEVIS system to a file format that UniPlot imports (see Importing Data).
15.5.9. Importing an Excel file in the Norm-Name configuration creates a file named “0” or “1” and no channels are translated¶
There are two possible causes:
No name was entered for the Norm-Name table. Enter a path and a file name for the Norm-Name table (
.unt) in the Norm-Name configuration dialog before you import the Excel file. If the field is empty, only a file whose name is a number is created and no translation is applied.In releases before R2020.5 the Norm-Name dialog had an error. Symptoms are the success message “0 created”, the table file name “0” or “1” and no
.untfile. The error is corrected in R2020.5. After the update delete the table name “0” or “1” once, enter a valid file name and import the Excel file again.
A formula list that cannot be imported can also be caused by missing write permission for the UniPlot user directory (see Help=>Open User Directory).
See also: Norm-Names.
15.6. Emission and RDE Add-ins¶
15.6.1. Which emission evaluation add-ins exist and which standards do they support?¶
There are two separate add-ins:
The Emission Evaluation add-in (
rs_emission, menu Emission Evaluation) evaluates the tests listed in Emission Evaluation (for example WHTC, WHSC, ETC, ESC, NRTC, NRSC, WNTE, FTP HD, JE05, ELR and EPA 1065).The Real Driving Emission (RDE) Data Evaluation add-in evaluates PEMS data of on-road tests and calculates conformity factors, see Real Driving Emission Data Evaluation.
The Emission Evaluation add-in is not part of the standard installation. The installation steps are described in Installation.
15.6.2. Where do I find the RDE (ISC) evaluation and its help?¶
RDE (Real Driving Emissions) and ISC (In-Service Conformity) are two names for the same add-in.
Enable the add-in Real-Driving Emissions - RDE Data Evaluation (BETA 002) in Tools=>Add-In Manager.
Choose RDE Evaluation > Evaluation… (see Usage). The evaluation window opens.
Click Help in this window. The manual that opens is titled “In-Service Conformity Evaluation”. This is the help for the RDE evaluation.
See also: Real Driving Emission Data Evaluation, Usage.
15.6.3. Where do I set the emission limits in the RDE (Real Driving Emissions) add-in?¶
You cannot change the emission limits in a dialog. The add-in uses fixed WHTC Euro VI (diesel) limit values, for example NOx 0.46 g/kWh, CO 4.0 g/kWh and THC 0.16 g/kWh. The limit used for each component is written into the result file as the attribute e.limit_<component>.
The menu RDE Evaluation has two commands for settings:
RDE Evaluation > Configuration… sets only the directory for the result files.
RDE Evaluation > Evaluation… sets the evaluation parameters: Window Method, WHTC CO2-Mass in kg, WHTC Work in kWh, Max. Power in kW, Perform Dry-to-Wet Correction and Apply Gas-Response Time. You can save these settings as a named Configuration.
Other limits need an adapted add-in script (rs_pems_create.ic in <uniplot>\addin\rs_pems). Ask UniPlot support for help.
See also: Real Driving Emission Data Evaluation, Usage.
15.6.4. The RDE add-in shows different values on the first page and on later pages when I evaluate the same NC file twice¶
Run the add-in on a freshly created NC file: import the measurement again before you evaluate again.
The cause: each run stores its results in the global attributes of the NC file. The first page of the template reads its values from these attributes, which can be old values. Later pages use the current result from the memory of the add-in. The attribute names of saved computed channels also get numbered (_1, _2, …).
15.6.5. The particle number (PN) conformity factor of the RDE add-in is much lower than in another tool¶
In versions before R2021.3 the calculation of the particle number conformity factor used the PN information in g/s instead of #/s, which underestimated the conformity factor. The error is corrected in R2021.3. Update UniPlot to R2021.3 or later and repeat the evaluation.
15.6.6. The emission add-in writes a log error about a missing icon file soot.bmp at every start¶
Some packages of the Emission Evaluation add-in do not contain the icon file soot.bmp. The message is harmless. To avoid it, copy an icon with that name into the add-in folder addin\rs_emission.
See also: Installation.